DIAL-278: implementation of heat disinfection state timer overrides.

  1. … 3 more files in changeset.
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16309 added a new heat disinfect state

  1. … 1 more file in changeset.
Done.

Done.

Done.

Done.

DEN-16093 code review comments added

DEN-16093 fixed code review comments

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is redu...

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is reduced, this level should also be reduced accordingly."

100% now.

100% now.

DEN-16093 fixed merging conflicts

  1. … 3 more files in changeset.
DEN-16093-Increased conductivity percentage check and time out values

take latest changes from staging

  1. … 3 more files in changeset.
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DIAL-276_DG Event Event ID 15 Prints Duplicate Events With Different State Time
DG-DIAL-276_DG Event Event ID 15 Prints Duplicate Events With Different State Time
DIAL-276: fixed op status event duplicating on op mode change only.

Merge remote-tracking branch 'origin/DEN-16280-stuck-in-post-treatment-when-inlet-water-alarm-appears' into develop

  1. … 5 more files in changeset.
DEN-16093-merge in to develop

  1. … 1 more file in changeset.
DEN-16093-changed prime Max time and conductivity poercent check

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.