Fixed. Thanks!

Fixed. Thanks!

Please add _MS to be consistent with the reset of times which are all in milliseconds.

Please add _MS to be consistent with the reset of times which are all in milliseconds.

DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Done.

Done.

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is redu...

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is reduced, this level should also be reduced accordingly."

100% now.

100% now.

DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DIAL-276_DG Event Event ID 15 Prints Duplicate Events With Different State Time
DG-DIAL-276_DG Event Event ID 15 Prints Duplicate Events With Different State Time
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Add a blank line between declarations and code.

Add a blank line between declarations and code.

#includes are sorted. Put OperationModes.h in its proper place.

#includes are sorted. Put OperationModes.h in its proper place.

required changes are done.

required changes are done.

Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting...

Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting). Function would also create and send the cmd response message and return TRUE/FALSE which would set local result here for the ack response message below.
We should reject cmd if pumps are busy/on and/or if current DG mode is fill or fault mode.

DG-DEN-16093_Alarm 97 Conductivity Out OF Range Triggered ON First Fill OF Treatment
DG-DEN-16093_Alarm 97 Conductivity Out OF Range Triggered ON First Fill OF Treatment
DG-DEN-16250_Flow Check Alarm 207 When Filling With Pause State
DG-DEN-16250_Flow Check Alarm 207 When Filling With Pause State