DEN-16323, DEN-16277 updated the chemical disinfect to get the level of the reservoir under disinfection more accurately

DG-DEN-15791_DG Software Design Description
DG-DEN-15791_DG Software Design Description
DEN-16309 updated the code. Changed the load cells' full level from 1900 to 1850 mL in heat disinfect to be consistent with the rest of the cleaning modes

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DIAL-281: Code review comments.

Fixed. Thanks!

Fixed. Thanks!

Please add _MS to be consistent with the reset of times which are all in milliseconds.

Please add _MS to be consistent with the reset of times which are all in milliseconds.

DIAL-278: implementation of heat disinfection state timer overrides.

  1. … 3 more files in changeset.
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16309 added a new heat disinfect state

  1. … 1 more file in changeset.
Done.

Done.

Done.

Done.

DEN-16093 code review comments added

DEN-16093 fixed code review comments

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is redu...

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is reduced, this level should also be reduced accordingly."

100% now.

100% now.

DEN-16093 fixed merging conflicts

  1. … 3 more files in changeset.
DEN-16093-Increased conductivity percentage check and time out values

take latest changes from staging

  1. … 3 more files in changeset.
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT