Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Done.

Done.

Why 0.0F? memset() doesn't care about type - it just sets a number of bytes to given value.

Why 0.0F? memset() doesn't care about type - it just sets a number of bytes to given value.

DG-DEN-15791_DG Software Design Description
DG-DEN-15791_DG Software Design Description
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Please add _MS to be consistent with the reset of times which are all in milliseconds.

Please add _MS to be consistent with the reset of times which are all in milliseconds.

DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Done.

Done.

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is redu...

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is reduced, this level should also be reduced accordingly."

100% now.

100% now.

DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DIAL-276_DG Event Event ID 15 Prints Duplicate Events With Different State Time
DG-DIAL-276_DG Event Event ID 15 Prints Duplicate Events With Different State Time
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.