DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Why 0.0F? memset() doesn't care about type - it just sets a number of bytes to given value.

Why 0.0F? memset() doesn't care about type - it just sets a number of bytes to given value.

DG-DEN-15791_DG Software Design Description
DG-DEN-15791_DG Software Design Description
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Please add _MS to be consistent with the reset of times which are all in milliseconds.

Please add _MS to be consistent with the reset of times which are all in milliseconds.

DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Done.

Done.

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is redu...

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is reduced, this level should also be reduced accordingly."

100% now.

100% now.

DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT