DEN-16323 added code to reset the conductivity and temperature out of range persistent time in chemical disinfect

DG-DEN-16323_FW Alarm 243 Disinfection Interrupted Triggers During Chemical Disinfect
DG-DEN-16323_FW Alarm 243 Disinfection Interrupted Triggers During Chemical Disinfect
DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Initial changes for alarm 206 timer

  1. … 1 more file in changeset.
DG-DEN-16314_Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
DG-DEN-16314_Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15614 updated the chemical disinfect VPd to drain in the last drains, DEN-16354 changed the heat disinfect 500 mL timeout to 2 minutes to align with the reqruirements

DEN-15614 updated the chemical disinfect VPd to drain in the last drains

DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-16309 merged staging

DEN-16309 addressed a code review comment

Done.

Done.

Why 0.0F? memset() doesn't care about type - it just sets a number of bytes to given value.

Why 0.0F? memset() doesn't care about type - it just sets a number of bytes to given value.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
Merged DEN-16309

  1. … 2 more files in changeset.
DEN-16323, DEN-16277 updated the chemical disinfect to get the level of the reservoir under disinfection more accurately

DG-DEN-15791_DG Software Design Description
DG-DEN-15791_DG Software Design Description
DEN-16309 updated the code. Changed the load cells' full level from 1900 to 1850 mL in heat disinfect to be consistent with the rest of the cleaning modes

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.