DEN-16383 updated the code

These fields have been moved to a different broadcast at some point, and were no longer being set by the reservoir data broadcast. The fields regularly contained "NAN" values, so I removed them.

These fields have been moved to a different broadcast at some point, and were no longer being set by the reservoir data broadcast. The fields regularly contained "NAN" values, so I removed them.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-16383 updated mode flush

Why are these fields removed?

Why are these fields removed?

DEN-16383 fixed the heat disinfect alarm 242 trigger

DG-DEN-16383_Heat Disinfect R1 TO R2 Timer Overflows And Stuck IN State
DG-DEN-16383_Heat Disinfect R1 TO R2 Timer Overflows And Stuck IN State
Merge remote-tracking branch 'origin/DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low' into develop

Done. Code updates pushed.

Done. Code updates pushed.

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Updated headers

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

fill time allowance for time spent in pause state

Cleared fill timer on transition to generate

Added removed unused (and uninitialized) variables from broadcast data structure.

  1. … 1 more file in changeset.
I don't see these being used anywhere.

I don't see these being used anywhere.

DG Staging Unreviewed Commits
DG Staging Unreviewed Commits
DG-DEN-16381_Chemical Disinfect 1St Transition TO Disinfect R1 TO R2 Lasting Only 50 MS
DG-DEN-16381_Chemical Disinfect 1St Transition TO Disinfect R1 TO R2 Lasting Only 50 MS
DEN-16381 updated the doxygen comment

DEN-16381 added code to 0 the disinfect timer prior to transitioning to disinfect R1 to R2 for the first time

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-16323

  1. … 7 more files in changeset.
DEN-16323 added code to reset the conductivity and temperature out of range persistent time in chemical disinfect

DG-DEN-16323_FW Alarm 243 Disinfection Interrupted Triggers During Chemical Disinfect
DG-DEN-16323_FW Alarm 243 Disinfection Interrupted Triggers During Chemical Disinfect
DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Initial changes for alarm 206 timer

  1. … 1 more file in changeset.
DG-DEN-16314_Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
DG-DEN-16314_Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces