Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
These to values appear to be minimums. Why are they being called maximums?

These to values appear to be minimums. Why are they being called maximums?

DEN-15973 fixed the inlet water pressure alarm. DEN-16418 added chemical flush done after chemical disinfect prior to starting RO permeate sample mode

  1. … 1 more file in changeset.
DEN-15973 added the inlet water pressure limit in flush drain in the cleaning modes

  1. … 1 more file in changeset.
Merge remote-tracking branch 'origin/DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low' into develop

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

On Dara's advice, ifdef'ed code for fill state treatment recovery

DG-DEN-15973_Alarm 48 UF Rate Tare Error Triggered During Treatment
DG-DEN-15973_Alarm 48 UF Rate Tare Error Triggered During Treatment
Merge remote-tracking branch 'origin/DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low' into develop

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Reset timers when doing a test recover state, or returning from a (reset) fault condition.

DIAL-283: compiler fixes.

DIAL-283: message handling for moving average override.

  1. … 3 more files in changeset.
DIAL-283: chem disinfect moving average override

DG-DIAL-283_NO Moving Average Commands Conductivity
DG-DIAL-283_NO Moving Average Commands Conductivity
Merge remote-tracking branch 'origin/DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low' into develop

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Reintroduced reservoir data structure change.

Corrected the setModeFillStateTransition semantics.

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Corrected spacing in the file per CR comments

Add blank line between this code and switch below.

Add blank line between this code and switch below.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merge remote-tracking branch 'origin/DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low' into develop

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Follow up for changes, simplifications suggested by Sean.