Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15973 addressed the code review comments

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Dara, I mean inline comments for the magic numbers. Explain what 40 and 32 mean and why you're using those values.

Dara, I mean inline comments for the magic numbers. Explain what 40 and 32 mean and why you're using those values.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-15973 addressed the code review comments

  1. … 1 more file in changeset.
Done

Done

Done

Done

You at least need to add a comment explaining what they are.

You at least need to add a comment explaining what they are.

Done

Done

No, these are the minimum values to trigger the alarm. But they are named maximums.

No, these are the minimum values to trigger the alarm. But they are named maximums.

Yes these are for software configuration disable only so I did not make any #defines for them.

Yes these are for software configuration disable only so I did not make any #defines for them.

Yes, so name should reflect that. These names suggest they are minimum pressures below which we would alarm.

Yes, so name should reflect that. These names suggest they are minimum pressures below which we would alarm.

Align comment.

Align comment.

Magic numbers?

Magic numbers?

This is the minimum values that are needed to clear the alarm.

This is the minimum values that are needed to clear the alarm.

This is the minimum values that are needed to clear the alarm.

This is the minimum values that are needed to clear the alarm.

These 2 values appear to be minimum recovery thresholds.

These 2 values appear to be minimum recovery thresholds.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
These to values appear to be minimums. Why are they being called maximums?

These to values appear to be minimums. Why are they being called maximums?

DEN-15973 fixed the inlet water pressure alarm. DEN-16418 added chemical flush done after chemical disinfect prior to starting RO permeate sample mode

  1. … 1 more file in changeset.
DEN-15973 added the inlet water pressure limit in flush drain in the cleaning modes

  1. … 1 more file in changeset.
Merge remote-tracking branch 'origin/DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low' into develop

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

On Dara's advice, ifdef'ed code for fill state treatment recovery