Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-3421 work in progress

  1. … 2 more files in changeset.
DEN-4184: Added alarms when inlet water pressure low

  1. … 2 more files in changeset.
DEN-4217: Added and implemented accelerometer module. Increased fan PWM frequency to 25KHz.

  1. … 14 more files in changeset.
DEN-3421 Work in progress, added code to heat disinfect states

  1. … 1 more file in changeset.
DEN 4184: Added self test for pressure sensors

  1. … 2 more files in changeset.
DEN-3421 added functions to DrainPump.c/h and ROPump.c/h to support open loop and closed loop control modes. Added the command functions for Dialin

  1. … 7 more files in changeset.
DEN-4169: Check inlet water temperature in Fill, Drain, and Re-circulate modes

  1. … 2 more files in changeset.
DEN-3421 added more functions to RO pump

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 22 more files in changeset.
Finish fix to Valves module. Allow transition to heat disinfect mode from standby mode. Fix CAN comm logic for when DG is alone on bus.

  1. … 7 more files in changeset.
DEN-3421 modified the RO pump module and added commands to run from SystemComm

  1. … 12 more files in changeset.
Removed V1 build option. UT updates.

  1. … 5 more files in changeset.
DEN-3922 Addressed code reviews and removed VC previous import errors

  1. … 23 more files in changeset.
DEN-3922 Addressed code reviews and removed VC previous import errors

  1. … 5 more files in changeset.
Merge branch 'DEN-2490-firmware-lockup' into DEN-3421-dg-heat-disinfection

  1. … 7 more files in changeset.
Resolved a merge conflict with master

  1. … 10 more files in changeset.
Handle CAN and UART error/warning notifications. Differentiated HD from DG s/w fault. Updated some older module/function headers to latest format.

  1. … 22 more files in changeset.
DEN-3922 Fixed doxygen errors and corrected spelling errors

  1. … 5 more files in changeset.
DEN-3922 Check inlet water conductivity during Fill, Drain, and Recirculate Modes

  1. … 4 more files in changeset.
Fixed bug where new debug event for full buffer would overflow a temporary buffer to hold the event text. Removed some obsolete comm buffer protection code. Removed obsolete CAN test build option.

  1. … 10 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 60 more files in changeset.
UT and IT updates. Added IT for Reservoirs.c.

  1. … 1 more file in changeset.
DEN-3421 modified RO pump and Drain pump and added a control loop to them.

  1. … 5 more files in changeset.
Fix reservoir mix-up in Drain/Fill modes. Added test code to turn fan on when starting re-circ mode to cool heater until fan module is written.

  1. … 1 more file in changeset.
UT updates. Added UT for Reservoirs module. Minor fixes per UT.

  1. … 1 more file in changeset.
DEN-3421 work in progress. Added a state to ramp up the RO pump for the targeted flow

  1. … 9 more files in changeset.
UT updates. Updated RO pump PI coefficients to make less aggressive based on lab testing.

  1. … 4 more files in changeset.
Changed state of VRc from re-circ state to drain state in re-circulate mode per latest SA.

DEN-3421 added functions and states to DG heat disinfect mode and initial release of the UVReactors

  1. … 2 more files in changeset.
More EMC test build changes. Set Primary heater PWM freq to 15KHz.