Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-16093 code review comments added

DEN-16093 fixed code review comments

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is redu...

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is reduced, this level should also be reduced accordingly."

100% now.

100% now.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-16093 fixed merging conflicts

DEN-16093-Increased conductivity percentage check and time out values

take latest changes from staging

As we discussed, I think alarm 150 and related inlet water alarms are already set to block triggering during rinseback and post-treatment modes. So these changes are likely not needed.

As we discussed, I think alarm 150 and related inlet water alarms are already set to block triggering during rinseback and post-treatment modes. So these changes are likely not needed.

DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

This was a specific change for this alarm (150), and the associated alarms in this routine. It appeared there were similar mode limits for temperature as well. Pressure limits were more uniquely re...

This was a specific change for this alarm (150), and the associated alarms in this routine. It appeared there were similar mode limits for temperature as well. Pressure limits were more uniquely related to valve position, I thought.

I had seen the related changes in 59, and included them in my develop build.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

We don't trigger inlet water conductivity alarm in post-treatment mode? Any reason why we're not just saying "== MODE_POST"? What about inlet water pressure/temperature? FYI - getHDOperationMode()...

We don't trigger inlet water conductivity alarm in post-treatment mode? Any reason why we're not just saying "== MODE_POST"? What about inlet water pressure/temperature?

FYI - getHDOperationMode() was not working - HD op mode handler was mishandling message - fixed in my DENBUG-59 branch, but not yet in staging.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged staging to DENBUG-59.

DEN-16091 changed the trimmer heater's persistent alarm from 10 seconds to 12 seconds

DG-DEN-16308_Srsdg 53 Not Met
DG-DEN-16308_Srsdg 53 Not Met