Add blank line between this code and switch below.

Add blank line between this code and switch below.

Keep declarations at top of scope. Just have assignments down here.

Keep declarations at top of scope. Just have assignments down here.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

These fields have been moved to a different broadcast at some point, and were no longer being set by the reservoir data broadcast. The fields regularly contained "NAN" values, so I removed them.

These fields have been moved to a different broadcast at some point, and were no longer being set by the reservoir data broadcast. The fields regularly contained "NAN" values, so I removed them.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Why are these fields removed?

Why are these fields removed?

DG-DEN-16383_Heat Disinfect R1 TO R2 Timer Overflows And Stuck IN State
DG-DEN-16383_Heat Disinfect R1 TO R2 Timer Overflows And Stuck IN State
Done. Code updates pushed.

Done. Code updates pushed.

I don't see these being used anywhere.

I don't see these being used anywhere.

DG Staging Unreviewed Commits
DG Staging Unreviewed Commits
DG-DEN-16381_Chemical Disinfect 1St Transition TO Disinfect R1 TO R2 Lasting Only 50 MS
DG-DEN-16381_Chemical Disinfect 1St Transition TO Disinfect R1 TO R2 Lasting Only 50 MS
DG-DEN-16323_FW Alarm 243 Disinfection Interrupted Triggers During Chemical Disinfect
DG-DEN-16323_FW Alarm 243 Disinfection Interrupted Triggers During Chemical Disinfect
Is the "0" an ACK,NACK, PENDING_OVERRIDE or some other flag? Suggest a #define. (or a comment)

Is the "0" an ACK,NACK, PENDING_OVERRIDE or some other flag? Suggest a #define. (or a comment)

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

We typically dont #define for 0. However, the logic has been reworked that we no longer need to compare by 0.

We typically dont #define for 0. However, the logic has been reworked that we no longer need to compare by 0.

DG-DEN-16314_Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
DG-DEN-16314_Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List