Merged DEN-11114 and resolved merge conflict

DEN-11114 added header to the functions in FPGA

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2

DEN-11114: Removed arterial bubble sensor.

DEN-11114 worked on the reservoirs and blood leak detector

    • -19
    • +75
    /firmware/App/Controllers/BloodLeak.c
Merge branch 'DEN-11098-implement-hd-pre-treatment-mode-update' into develop

Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

DEN-11098 Converted from seconds to milliseconds fix

Fixed merge conflict

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

Enabled the skip reservoirs alarm

Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

DEN-11114: Fixed mid-treatment changes not immediately sending new ranges to UI problem.

DEN-11098 Corrected state transition and changed blood pump speed from 400 to 300 per system specification

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2

DEN-11114: ifdef'd dead code in alarm lamp driver. Clear POR bit after logging reset in POST.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-10602_UI_DEV_Add_PatientID_PreTreatment' into develop

    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.c
DEN-11543: Unable proceed to the Confirm Treatment when "08-3251-9" Acid Concentrate is selected

The HD FW updated to accept up to 4 acid concentration type index.

Below is the UI DataList.conf for acid concentrate.

[Acid Concentrate]

Fres. Naturalyte

08-1251-1

08-2251-0

08-3251-9

    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.c
DEN-11098 Incorporated peer review comments

Yes, we could use a single variable as the start time in all the pre-treatment states

Yes, we could use a single variable as the start time in all the pre-treatment states

Replaced "loadcellSteadyVolumeStartTime" with "steadyVolumeSamplingStartTime"

Replaced "loadcellSteadyVolumeStartTime" with "steadyVolumeSamplingStartTime"

Done

Done

Done

Done

No, loadcellSteadyVolumeStartTime was last set when the primeDialysateDialyzerTimeLimit has just expired.

No, loadcellSteadyVolumeStartTime was last set when the primeDialysateDialyzerTimeLimit has just expired.

Done

Done

Done

Done

Done.

Done.