What if we got some other alarm during pressure test states? Would it be better to check dryPressureTestsCompleted?

What if we got some other alarm during pressure test states? Would it be better to check dryPressureTestsCompleted?

Put constant on other side of comparison operator. We only put constant first when using "==".

Put constant on other side of comparison operator. We only put constant first when using "==".

This if will prevent countdown timer from being displayed unless we are paused or stopped after rinseback. I think countdown timer is definitely needed for rinseback init state (pump is stopped and...

This if will prevent countdown timer from being displayed unless we are paused or stopped after rinseback. I think countdown timer is definitely needed for rinseback init state (pump is stopped and rinseback hasn't even started). May be needed for new reconnect patient state as well.
I think it might be simpler/clearer to replace this if with a check for whether BP is running or not. If BP is stopped, we proceed to next if. If BP is running, keep timeouts at zero.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

HD-DEN-16164_FW Pre Treatment Pressure Test Failures Vent TO Saline Bag
HD-DEN-16164_FW Pre Treatment Pressure Test Failures Vent TO Saline Bag
HD-DEN-16239_Add A Test Configuration TO Disable Blood Leak Alarm
HD-DEN-16239_Add A Test Configuration TO Disable Blood Leak Alarm
Done.

Done.

Add syringePumpStateStartTime to header.

Add syringePumpStateStartTime to header.

HD-DEN-14692_Need Override For Alarm ID HD Cpu Ram Error
HD-DEN-14692_Need Override For Alarm ID HD Cpu Ram Error
HD-DEN-16200_Syringe Pump Stuck IN Priming State
HD-DEN-16200_Syringe Pump Stuck IN Priming State
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

HD Staging Unreviewed Commits
HD Staging Unreviewed Commits