RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Yes. The other alarms are handled in different drivers or modes. The above alarms are here to make sure HD sends the fill command after the bottles have been replaced.

Yes. The other alarms are handled in different drivers or modes. The above alarms are here to make sure HD sends the fill command after the bottles have been replaced.

HD-DEN-16397_After Treatment Duration IS Changed Dpo Speeds UP And Alarm 127 Venous Pressure Low
HD-DEN-16397_After Treatment Duration IS Changed Dpo Speeds UP And Alarm 127 Venous Pressure Low
HD-DEN-16428_Need A Way OR Approach TO Validate The Srshd
HD-DEN-16428_Need A Way OR Approach TO Validate The Srshd
Just these 2 alarms? Should we check for any alarm that STOPS treatment?

Just these 2 alarms? Should we check for any alarm that STOPS treatment?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Done

Done

Done

Done

I don't think we want to use recorded volume (would want inactive reservoir if we did). I think we want to use your function here that calculates the target fill volume.

I don't think we want to use recorded volume (would want inactive reservoir if we did). I think we want to use your function here that calculates the target fill volume.

Should we have a local var for active reservoir too?

Should we have a local var for active reservoir too?

Should this be active reservoir?

Should this be active reservoir?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Done

Done

General issue - please consider appropriateness of using this calculation when calling this function.

General issue - please consider appropriateness of using this calculation when calling this function.

I think we have to remember last fill volume for each reservoir and refer to that. With this approach, if user changes Qd, the percentage will be wrong.

I think we have to remember last fill volume for each reservoir and refer to that.
With this approach, if user changes Qd, the percentage will be wrong.

HD-DEN-16442_Reservoir Fill Change For Clearance Performance Improvement
HD-DEN-16442_Reservoir Fill Change For Clearance Performance Improvement