Thanks. Was not sure about that but wanted to cover it if bolus extended to treatment mode. Removed.

Thanks. Was not sure about that but wanted to cover it if bolus extended to treatment mode. Removed.

HD-DEN-16372_Srsdg 907 Not Met
HD-DEN-16372_Srsdg 907 Not Met
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

We should not be doing a Heparin bolus in treatment mode. It should be completed in pre-treatment mode. transitionToDialysis() function will resume Heparin dispense (not bolus) if necessary when tr...

We should not be doing a Heparin bolus in treatment mode. It should be completed in pre-treatment mode. transitionToDialysis() function will resume Heparin dispense (not bolus) if necessary when treatment is resumed.

Are these F32s?

Are these F32s?

HD-DENBUG-84_Heparin Bolus Does Not Resume After Treatment Pause
HD-DENBUG-84_Heparin Bolus Does Not Resume After Treatment Pause
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

HD-DEN-16090_2 Sys Saline Bubbles Enter Pbo Line During Cartridge Priming
HD-DEN-16090_2 Sys Saline Bubbles Enter Pbo Line During Cartridge Priming
HD-DENBUG-83_Alarm 278 Triggered Active Res Weight Out OF Range
HD-DENBUG-83_Alarm 278 Triggered Active Res Weight Out OF Range