DEN-11098 Added compiler switch RUN_BLOOD_CIRCUIT_PRIME_AGAIN

DEN-11114 continued working on the reservoir management

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -10
    • +3
    /firmware/App/Services/Reservoirs.c
DEN-11089 Added cases in switch statement to handle prime pause for new saline dialyzer state

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2

DEN-11114 continued working on the resevoir management

    • -7
    • +49
    /firmware/App/Controllers/DGInterface.c
    • -2
    • +15
    /firmware/App/Controllers/DGInterface.h
    • -1
    • +1
    /firmware/App/Modes/ConsumableSelfTest.c
    • -0
    • +3
    /firmware/App/Modes/ModeTreatmentParams.c
    • -42
    • +119
    /firmware/App/Services/Reservoirs.c
DEN-11114: perform periodic clock check only if release build so debugging will not cause alarm.

DEN-11089 Removed signalDialOutPumpHardStop since it has not been started

DEN-11114: Removed obsolete UF PI controller. Cleaned up UF control code changes.

    • -19
    • +15
    /firmware/App/Controllers/DialOutFlow.c
    • -1
    • +0
    /firmware/App/Services/PIControllers.c
    • -2
    • +1
    /firmware/App/Services/PIControllers.h
DEN-11089 the VDO pump is stopped before transition from dialyzer state to wet selftests state

DEN-11089 Correction - dialyzer dialysate bypass priming path uses R1 instead of R2 and dialyzer dialysate priming path uses R2 instead of R1

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2

DEN-11114: Added comment.

    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.c
Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2

    • -1
    • +16
    /firmware/App/Modes/ModeTreatment.c
DEN-11114 initial commit of the reservoirs module. Continued with implementing the reservoir management

    • -19
    • +120
    /firmware/App/Controllers/DGInterface.c
    • -3
    • +4
    /firmware/App/Controllers/DGInterface.h
    • -0
    • +407
    /firmware/App/Services/Reservoirs.c
    • -0
    • +38
    /firmware/App/Services/Reservoirs.h
DEN-11114: Updated UF control to included estimated UF flow as input.

    • -4
    • +5
    /firmware/App/Controllers/DialInFlow.c
    • -40
    • +106
    /firmware/App/Controllers/DialOutFlow.c
    • -0
    • +1
    /firmware/App/Controllers/DialOutFlow.h
    • -1
    • +16
    /firmware/App/Modes/ModeTreatment.c
DEN-11089 Incorporated peer review comments

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.c
DEN-11114: Restored previous motor controller speed feedback conversion to RPM - it was right before.

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

    • -6
    • +8
    /firmware/App/Controllers/DialInFlow.c
    • -4
    • +4
    /firmware/App/Controllers/DialOutFlow.c
DEN-11114: Fixed conversion issue for pump speed ADC voltage levels.

    • -6
    • +8
    /firmware/App/Controllers/DialInFlow.c
    • -4
    • +4
    /firmware/App/Controllers/DialOutFlow.c
DEN-11114: Changed syringe ID measurement to median between min and max ID given by manufacturer. Included prime volume in calculation of volume needed to meet needs of treatment.

    • -2
    • +2
    /firmware/App/Controllers/SyringePump.c
DEN-11089 if priming for the 1st time, transition to the HD_PRIME_SALINE_SETUP_STATE

DEN-11089 In hadlePrimeDialysateBypassState, state should be HD_PRIME_RESERVOIR_ONE_FILL_COMPLETE_STATE

DEN-11089 Swapped the dialysate priming circuit - prime the R2 circuit first then prime the R1 circuit

DEN-11089 Implemented new saline dialyzer priming state to purge air from the dialyzer path

DEN-11114: Updated syringe inner diameter to improve dispense accuracy. Updated minimum dispense rate to 0.2 mL/hr.

    • -2
    • +2
    /firmware/App/Controllers/SyringePump.c
DEN-11114: Add 5 sec persistence to blood leak alarm. Make insufficient Heparin volume alarm pre-emptive during seek operation.

    • -7
    • +21
    /firmware/App/Controllers/BloodLeak.c
    • -11
    • +10
    /firmware/App/Controllers/SyringePump.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo