hdfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-5381: Moved DG start and valves homing into pre-treatment mode

DEN-5674: Fixed alarm audio volume control per HDD change.

DEN-5831: Removed eval board and unused includes

DEN-5381: Moved new alarms to available slots

DEN-5381: Increased filling volume to 1700 mL

DEN-5381: Updated hd to dg drain cmd

    • -6
    • +11
    /firmware/App/Controllers/DGInterface.c
    • -1
    • +8
    /firmware/App/Controllers/DGInterface.h
DEN-5381: Removed reservoir drain and fill from standby mode

DEN-5674: Pass valves check when disable valves build switch is on.

DEN-5674: Added handling for resume after alarm. Removed obsolete build switches.

    • -7
    • +2
    /firmware/App/Controllers/DialInFlow.c
    • -7
    • +2
    /firmware/App/Controllers/DialOutFlow.c
    • -20
    • +1
    /firmware/App/Modes/ModePostTreat.c
    • -29
    • +4
    /firmware/App/Modes/ModeTreatment.c
    • -21
    • +0
    /firmware/App/Modes/ModeTreatmentParams.c
  1. … 5 more files in changeset.
DEN-5674: Fix alarm state bug.

DEN-5674: Added alarm silence handling.

DEN-5674: Moved alarm lamp self test from early to later in sequence.

DEN-5674: temporary set alarm volume low for testing.

DEN-5674: Fixed alarms state update.

DEN-5674: Prevent resume/rinseback options if top alarm condition not yet cleared.

DEN-5381: Added reservoir management to prime sub-mode

DEN-5674: Completed UF volume setting pre-treatment.

    • -2
    • +18
    /firmware/App/Modes/ModeTreatmentParams.c
DEN-5674: Removed obsolete test code. Added new UF param msg handler.

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.c
    • -0
    • +14
    /firmware/App/Modes/ModeTreatmentParams.c
    • -3
    • +5
    /firmware/App/Modes/ModeTreatmentParams.h
  1. … 3 more files in changeset.
DEN-5381: Refactor alarm management module

    • -170
    • +27
    /firmware/App/Services/AlarmMgmt.c
DEN-5381: Added time out for purge air state

DEN-5674: Fixed valves POST call. Added alarm for FPGA comm timeout.

DEN-5381: Added handler for prime sub-mode states

DEN-5674: Removed alarm dialyzer bypass flag. Added alarm user option for alarms that simply require acknowledgement. Added some alarm action handling for treatment mode.

    • -0
    • +62
    /firmware/App/Modes/ModeTreatment.c
    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.c
    • -82
    • +126
    /firmware/App/Services/AlarmMgmt.c
DEN-5381: Created skeleton for pre-treatment mode

    • -24
    • +104
    /firmware/App/Modes/ModePreTreat.c
    • -0
    • +54
    /firmware/App/Modes/Prime.c
    • -0
    • +44
    /firmware/App/Modes/Prime.h
DEN-5674: Added mode specific masking of enabled user alarm recovery actions.

    • -0
    • +5
    /firmware/App/Modes/ModeTreatmentParams.c
DEN-5674: Fixed drain reservoir cmd to DG to include indication of whether reservoir load cells should be tared.

    • -6
    • +55
    /firmware/App/Controllers/DGInterface.c
    • -2
    • +12
    /firmware/App/Controllers/DGInterface.h
DEN-5674: Added alarm (empty) action signal handling functions to the various op modes.

    • -1
    • +15
    /firmware/App/Modes/ModePostTreat.c
    • -0
    • +14
    /firmware/App/Modes/ModeTreatment.c
    • -1
    • +15
    /firmware/App/Modes/ModeTreatmentParams.c
  1. … 5 more files in changeset.
DEN-5674: Added alarm sub-rank property and use for determining top priority alarm for UI.

    • -21
    • +61
    /firmware/App/Services/AlarmMgmt.c
DEN-5674: Added synchronize of alarm audio/lamp control with published alarm state so that UI can flash alarm in sync w/ lamp. Added alarm audio.

    • -117
    • +74
    /firmware/App/Services/AlarmMgmt.c
DEN-5674: Updated alarm messaging. Added concept of alarm condition detected and ability to clear this. Clear all recoverable alarms when user chooses an action.

    • -172
    • +120
    /firmware/App/Services/AlarmMgmt.c