DialOutFlow.c

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
HD-DENBUG-104_Alarm 38 HD Fault Dpi Rotor T
HD-DENBUG-104_Alarm 38 HD Fault Dpi Rotor T
HD-DENBUG-104_Alarm 38 HD Fault Dpi Rotor T
HD-DENBUG-104_Alarm 38 HD Fault Dpi Rotor T
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Do we want to have the 11% minimum here? I thought it was just for closed loop control.

Do we want to have the 11% minimum here? I thought it was just for closed loop control.

This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit?

This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit?

HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate
HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Consider splitting this if and the setDialysisParams() function into two: 1) blood flow rate changed and 2) dialysate flow rate changed.

Consider splitting this if and the setDialysisParams() function into two: 1) blood flow rate changed and 2) dialysate flow rate changed.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.