HD-DENBUG-206_Venous Pressure Readings Were
HD-DENBUG-206_Venous Pressure Readings Were
HD-DENBUG-206_Venous Pressure Readings Were
HD-DENBUG-206_Venous Pressure Readings Were
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Since the utility is finding the difference of two numbers, I see the logic is common. for instance, On a positive case, u32DiffWithWrap(9000, 26000) where 9000 is the adjustment number and 26000 i...

Since the utility is finding the difference of two numbers, I see the logic is common. for instance, On a positive case, u32DiffWithWrap(9000, 26000) where 9000 is the adjustment number and 26000 is the current timer value, the result is 26000-9000 = 17000. on a negative case, u32DiffWithWrap(9000,4000) where current timer value is 4000, the result would be 0xFFFF FFFF - 9000 + 4000 + 1. i.e: FFFF FFFF - 5000 + 1. Do you still see any concern?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

This looks backwards and nonsensical to me. Does this work?

This looks backwards and nonsensical to me. Does this work?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Why is air pump on state checked here? Seems redundant as it is checked again below.

Why is air pump on state checked here? Seems redundant as it is checked again below.

The lower level sensor doesn't need to be raw here.