PresOccl.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 14 more files in changeset.
Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

DEN-13154: Removed obsolete pressure data from broadcast. Tightened no cartridge pressure range for PBA and PBo.

  1. … 1 more file in changeset.
Fixed merge conflicts.

  1. … 3 more files in changeset.
DEN-13154: Ensure alarm system has control of alarm lamp in fault mode in case fault occurred before lamp test completed. Do not request DG active alarms on startup if DG is not communicating.

  1. … 3 more files in changeset.
Merge branch 'DEN-14160-hd-blood-leak-override-not-working' into develop

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
Merge branch 'DEN-14589-DIAL-211' into develop

  1. … 5 more files in changeset.
Merge branch 'staging' into DEN-14589-DIAL-211

  1. … 4 more files in changeset.
Fixed merge conflicts.

  1. … 14 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 46 more files in changeset.
Fixed merge conflicts. Re-applied recently rolled back pre-tx state changes.

  1. … 15 more files in changeset.
DEN-14549: Fixed merge from staging merge conflicts.

  1. … 17 more files in changeset.
DEN-14556: updated alarm pressure occlusion alarm to 3s. HD now sends events when occlusion baseline is set. Alarm id 142 and 277 now send baseline occlusion value and trigger value.

DEN-14549: Added alarm threshold to occlusion alarm data for logging. Only check for sufficient Heparin when in pre-tx mode. Addressed code review comments.

  1. … 2 more files in changeset.
Merge branch 'DEN-14589-DEN-14011_fix' into develop

  1. … 3 more files in changeset.
DEN-14549: Update to arterial/venous pressure limit alarm logic. Update measured UF volume in all dialysis states. Remove dialysis and UF start states.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-14549: Fix art/ven pressure alarm check logic - apply user set limits in dialysis state (if not saline bolus) or treatment stop state (if BP running). Require door/latch to be closed in patient connection state. Remove escalation for prime complete alarms. Handle prime complete alarms in pre-tx recirculate state. Update prime complete alarm properties.

  1. … 6 more files in changeset.
DEN-14011: created handleDrySelfTestStartState. code review comments.

  1. … 1 more file in changeset.
DEN-14011: added condition to check if BP is running while we do occlusion checks.

DEN-14549: Rolled back change to include blood prime for normal inline pressure range checks. Leave as fixed max only range check in blood prime.

DEN-14549: include venous check in including blood prime.

DEN-14549: maintain blood flow rate when alarm w/ blood recirc property pauses treatment. maintain Heparin dispense when blood recirc is occurring during alarm pause of treatment. include blood prime state in checks for inline pressure range.

  1. … 3 more files in changeset.
DEN-14549: Misc updates to dry self-tests in pre-treatment.

  1. … 3 more files in changeset.
Fixed merge conflicts.

  1. … 17 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 10 more files in changeset.
Merge branch 'staging' into DEN-14344-sw-s85-mg-air-pump

  1. … 8 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 112 more files in changeset.
Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

  1. … 1 more file in changeset.