RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

How is this if different than the else above? I don't think we need a separate if/else for not detected debounce - everything can be done in the if/else above.

How is this if different than the else above? I don't think we need a separate if/else for not detected debounce - everything can be done in the if/else above.

These should be initialized to zero in init function.

These should be initialized to zero in init function.

DENBUG-326 added a lower debounce for the syringe pump detection voltage

We are only debouncing the transition from detected to not detected (syringe removal). Is that our design intent?

We are only debouncing the transition from detected to not detected (syringe removal). Is that our design intent?

DENBUG-326 added code to debounce the syringe detection voltage

HD-DENBUG-326_Syringe Switch Detection Failure Staging
HD-DENBUG-326_Syringe Switch Detection Failure Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
Merged DEN-17208

  1. … 16 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
DENBUG-232 resolving merge issues

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
DEN-16782 added the changes from staging

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DENBUG-216: Addressed potential issues with clearing of alarm persistence.

  1. … 1 more file in changeset.
HD-DENBUG-216_Alarm 89 Alarm ID DG Outlet U
HD-DENBUG-216_Alarm 89 Alarm ID DG Outlet U
HD-DEN-16782_Unable TO Clear Communication Error Alarm 65 During Power Loss
HD-DEN-16782_Unable TO Clear Communication Error Alarm 65 During Power Loss
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 18 more files in changeset.
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH