Fixed merge conflicts.

  1. … 2 more files in changeset.
DEN-7605: Added build switches.

  1. … 3 more files in changeset.
Need to remove this comment.

Need to remove this comment.

Need to remove this comment.

Need to remove this comment.

Same as comment from DialinFlow.c

Same as comment from DialinFlow.c

We can't have 3 calls to isPersistentAlarmTriggered() on same alarm. If the first call starts the timer and the second or third call is good, it will reset the timer to 0 again. Suggest creating a ...

We can't have 3 calls to isPersistentAlarmTriggered() on same alarm. If the first call starts the timer and the second or third call is good, it will reset the timer to 0 again.
Suggest creating a BOOL variable to have all three conditions or with each other and call isPersistentAlarmTriggered() only once.

DEN-7605: Fixed dialysate flow read count error typo.

v0.5.0.9035

  1. … 1 more file in changeset.
DEN-7860: Initial push for HD air bubble detectors module

  1. … 1 more file in changeset.
initBloodPrime() is called every time we transition to blood prime sub-mode.

initBloodPrime() is called every time we transition to blood prime sub-mode.

Blood prime last motor count does not seem to get updated except in initBloodPrime(). Should it be updated continuously somewhere?

Blood prime last motor count does not seem to get updated except in initBloodPrime(). Should it be updated continuously somewhere?

DEN-8103: Fixed build warnings

  1. … 9 more files in changeset.
DEN-8103: Cleaned up TODO and added alarms

  1. … 17 more files in changeset.
DEN-8103: Converted blood pump alarms to use persistent alarm

DEN-7605: Added FPGA read timeout checks/alarms for blood/dialysate flow.

  1. … 2 more files in changeset.
Fixed merge conflicts.

  1. … 3 more files in changeset.
DEN-7605: Added FPGA pressure sensor checks and alarms.

  1. … 2 more files in changeset.
DEN-7605: Updated peristaltic pump speed over PWM DC range for V3.

DEN-7605: Added FPGA reported comm error to FPGA comm error check. Fixed persistent alarm periods in syringe pump module.

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-7605: Added more published data and overrides to syringe pump module.

  1. … 4 more files in changeset.