DEN-13460 added the voltages persistent alarm

  1. … 1 more file in changeset.
DEN-13460: Added handling of request to re-send HD alarms. Expanded CAN message buffer sizes.

  1. … 6 more files in changeset.
DEN-13460: Add minimum syringe pump rate as criteria for syringe pump direction alarm.

DEN-13460: Use max time windowed count instead of persistence for commutation errors.

DEN-13460: Added minimum pump speed as criteria for pump direction alarms.

Updated the valves transition algorithm

DEN-13460 updated the pressure occlusion

  1. … 1 more file in changeset.
Updated the syringe pump retract function

  1. … 1 more file in changeset.
Merged DEN-13460 and resolved the conflicts

  1. … 1 more file in changeset.
Updated the drivers with the latest demo changes

  1. … 1 more file in changeset.
DEN-13460: Fixed syringe home issues.

  1. … 1 more file in changeset.
DEN-13460 updated the valves homing code

  1. … 2 more files in changeset.
DEN-13460: Relaxed alarm 159 threshold. Reverted back to old way of estimating implied motor speed from target flow for purposes of alarm 159.

DEN-13460: Fixed merge conflicts.

DEN-13460: Added persistence to pump direction error count alarm. Fixed comments and pos A average calc for valves.

  1. … 2 more files in changeset.
Merged DEN-13460

DEN-13460 updated the code for DVT

  1. … 2 more files in changeset.
Why removed?

Why removed?

Why removed? I thought we needed this for VC?

Why removed? I thought we needed this for VC?

F has been added.

F has been added.

No. Code removed.

No. Code removed.

If no actions handled in this mode, add a comment saying so.

If no actions handled in this mode, add a comment saying so.

Yes, at bottom of exec function above.

Yes, at bottom of exec function above.

Remove blank line.

Remove blank line.

Can we remove this now?

Can we remove this now?

Why?

Why?

Why?

Why?

HD-DEN-13141_Miscellaneous HD Issues BB
HD-DEN-13141_Miscellaneous HD Issues BB
DEN-13460: Fixed bug in syringe pump POST test.

DEN-13460: Baseline HD pump direction error counters after POST completes so we don't alarm due to power-up.