Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merged DEN-15367

  1. … 4 more files in changeset.
DEN-15367, DEN-15252, DEN-15484, DEN-15485 added code to process the cleaning modes if they were started directly from DG. Added code to be able to start RO permeate sample from UI settings. Added the blood leak detector counter to the broadcast

  1. … 3 more files in changeset.
Fixed merge conflict.

  1. … 1 more file in changeset.
DEN-15278: Added rotor hall sensor state to BP, DPi and DPo broadcast messages.

Add disinfectCancelReqID, currentStandbyState, currentDGCleaningMode

Add disinfectCancelReqID, currentStandbyState, currentDGCleaningMode

Isn't currentStandbyState an output?

Isn't currentStandbyState an output?

Add disinfectCancelReqID

Add disinfectCancelReqID

DEN-15404, DEN-15252, DEN-15485 work in progress. Implementing code to handle cleaning modes in HD standby mode and removed the old code

  1. … 2 more files in changeset.
HD-DEN-15278_Blood Stop Air Trap Alarm Declares During Saline Bolus Treatment
HD-DEN-15278_Blood Stop Air Trap Alarm Declares During Saline Bolus Treatment
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch 'DEN-15314-usability-update-alarm-text-to-not-use-acronyms' into develop

  1. … 1 more file in changeset.
DEN-15314: Added stall check to syringe prime operation. Extended dialysate bypass prime timeout and added a little histeresis to exit criteria for bypass prime state.

  1. … 1 more file in changeset.
HD-DEN-15314_Usability Update Alarm Text TO Not Use Acronyms
HD-DEN-15314_Usability Update Alarm Text TO Not Use Acronyms
Merged staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed merge conflicts from DEN-15347.

DEN-15347: Addressed issue recovering from venous occlusion alarm 208.

HD-DEN-15347_Usability Pressure Alarms During Recirculate Recirc State
HD-DEN-15347_Usability Pressure Alarms During Recirculate Recirc State
Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.