Merged staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-16385-treatment-log-event-ids-for-heparin-pause-are-not-notated-correctly' into develop

Merge branch 'DEN-16365-high-venous-alarm-does-not-appear-while-in-recirculation-mode' into develop

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16385 Send heparin stop/pause event when syringe pump stopped.

HD-DEN-16408_Dialysate Flow Performance Decrease AT 600Qd
HD-DEN-16408_Dialysate Flow Performance Decrease AT 600Qd
DEN-16408: Updated DPi and DPo modules to increase speed range to 0..4000 RPM in order to allow pumps to reach 600 mL/min rate consistently.

HD-DEN-16314_HD Den 16314 Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
HD-DEN-16314_HD Den 16314 Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
HD-DEN-16314-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Removed unused elements of Reservoir data.

HD-DEN-16314_Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
HD-DEN-16314_Alarm 206 Triggers After Resuming From Incoming Water Conductivity High OR Low
HD-DEN-16385_Treatment Log Event Ids For Heparin Pause Are Not Notated Correctly
HD-DEN-16385_Treatment Log Event Ids For Heparin Pause Are Not Notated Correctly
Merge remote-tracking branch 'origin/DEN-16314-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low' into develop

DEN-16314-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Updated the DG_RESERVOIR_DATA_T structure on the HD side

DEN-16385 Fix heparin pause command ID.

HD-DEN-16365_High Venous Alarm Does Not Appear While IN Recirculation Mode
HD-DEN-16365_High Venous Alarm Does Not Appear While IN Recirculation Mode
DEN-16365 Enable venous occlusion alarm in treatment recirculation mode. Other pressure alarms disabled.

Merged staging

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16200-2-merge in to staging