Discussed with Sean and confirmed that matching requirement done in the code. We are good with the modification now.

Discussed with Sean and confirmed that matching requirement done in the code. We are good with the modification now.

Why was the = removed from the ternary ifs?

Why was the = removed from the ternary ifs?

HD-DEN-16391_FW Dialysate Temperature Alarm Not Sensitive Enough Use Tdi Moving Avg
HD-DEN-16391_FW Dialysate Temperature Alarm Not Sensitive Enough Use Tdi Moving Avg
HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate
HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate
HD-DENBUG-157_2Alarm 48 UF Accuracy Triggers Immediately After Alarm 272 HD Fan Rpm Out OF Range
HD-DENBUG-157_2Alarm 48 UF Accuracy Triggers Immediately After Alarm 272 HD Fan Rpm Out OF Range
HD-DENBUG-158_2 Alarm 128 Venous Air Bubble Triggered Again During Post TX Sequence
HD-DENBUG-158_2 Alarm 128 Venous Air Bubble Triggered Again During Post TX Sequence
HD-DENBUG-198_2 Alarm 130 142 Triggered And The End OF A Rinseback End OF Treatment
HD-DENBUG-198_2 Alarm 130 142 Triggered And The End OF A Rinseback End OF Treatment
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.