Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Remove extra space after >. Add () around inner conditions.

Remove extra space after >. Add () around inner conditions.

Remove space before comma.

Remove space before comma.

Remove space before comma.

Remove space before comma.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-16563-venous-pressure-readings-were-below-20mmhg-for-1.5-minutes-before-low-venous

HD-DEN-16648_Update Syringe Pump Positioning IN Software Per Force Sensor Button Update
HD-DEN-16648_Update Syringe Pump Positioning IN Software Per Force Sensor Button Update
HD-DEN-16648_Update Syringe Pump Positioning IN Software Per Force Sensor Button Update
HD-DEN-16648_Update Syringe Pump Positioning IN Software Per Force Sensor Button Update
Merge branch 'DEN-16648-update-syringe-pump-positioning-in-software-per-force-sensor-button-update' into develop

DEN-16648: Changed home to empty volume from 11 to 10.755 mL due to extension of pusher puck length by 1.5 mm.

DEN-16648: Changed home to empty volume from 11 to 10.755 mL due to extension of pusher puck by 1.5 mm.

DENBUG-162 resolve merge issues

  1. … 1 more file in changeset.
DENBUG-162-2 pressure stabilzation changes

HD-DENBUG-162_Low Venous Pressure Alarm Triggers During Normal Treatment Operation
HD-DENBUG-162_Low Venous Pressure Alarm Triggers During Normal Treatment Operation
Done.

Done.

Done.

Done.

DENBUG-162 addressing code review comments

MIN_TIME_BETWEEN_AIR_TRAP_FILL_LOW_PRESSURE_ALARM_EXEMPTIONS_MS (as a suggestion). Also be more descriptive in comments.

MIN_TIME_BETWEEN_AIR_TRAP_FILL_LOW_PRESSURE_ALARM_EXEMPTIONS_MS (as a suggestion).
Also be more descriptive in comments.

Add OR condition for if useShort == FALSE

Add OR condition for if useShort == FALSE

Use u32DiffWithWrap() function in utilities.

Use u32DiffWithWrap() function in utilities.

This looks wrong.

This looks wrong.

DENBUG-162 pressure stabilization code changes

HD-DENBUG-162_Low Venous Pressure Alarm Triggers During Normal Treatment Operation
HD-DENBUG-162_Low Venous Pressure Alarm Triggers During Normal Treatment Operation
HD-DEN-16499_FW Message List Testing Issues
HD-DEN-16499_FW Message List Testing Issues