Use F32 macro since filtered pressure will be float.

Use F32 macro since filtered pressure will be float.

Must be in dialysis state and stage 2 or stable pressure state.

Must be in dialysis state and stage 2 or stable pressure state.

I think for partial occlusion, we need a separate 15 min timer and should not be a timeout - needs to be a timercounter so it can be paused when not in dialysis state.

I think for partial occlusion, we need a separate 15 min timer and should not be a timeout - needs to be a timercounter so it can be paused when not in dialysis state.

0.0F.

0.0F.

Should be a F32 since it's an average. Comment should say 10s or something generic.

Should be a F32 since it's an average. Comment should say 10s or something generic.

Not needed if you only trigger alarm after 1st stage destabilization state.

Not needed if you only trigger alarm after 1st stage destabilization state.

Change to "bloodPumpPartialOcclusionBaseline".

Change to "bloodPumpPartialOcclusionBaseline".

should be 1,000

should be 1,000

Capitalize Baseline.

Capitalize Baseline.

Change to "bldPumpOcclusionLongFilter".

Change to "bldPumpOcclusionLongFilter".

HD-DEN-16851_New Alarm For Increased Occlus
HD-DEN-16851_New Alarm For Increased Occlus
HD-DENBUG-239_Alarm 160 Blood Pump Rotor SP
HD-DENBUG-239_Alarm 160 Blood Pump Rotor SP
HD-DEN-16778_System Does Not Check That Air
HD-DEN-16778_System Does Not Check That Air
HD-DENBUG-240_After A Manual Rinseback And
HD-DENBUG-240_After A Manual Rinseback And
HD-DEN-16819_IN Treatment Parameter Mode NO Rejection Seen When Hep Rate OR Hep Bol Value IS Set TO Less Than 0
HD-DEN-16819_IN Treatment Parameter Mode NO Rejection Seen When Hep Rate OR Hep Bol Value IS Set TO Less Than 0
HD-DENBUG-232_Unable TO Clear Communication
HD-DENBUG-232_Unable TO Clear Communication