Param is wrong.

Param is wrong.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

HD-DENBUG-259_Low Venous Pressure Alarm Did
HD-DENBUG-259_Low Venous Pressure Alarm Did
This is just the active cool mode by itself.

This is just the active cool mode by itself.

We didn't have this before?

We didn't have this before?

HD-DEN-16949_Low Venous Pressure Alarm Did
HD-DEN-16949_Low Venous Pressure Alarm Did
HD-DEN-16680_P1B Inlet Water Shutdown During Heat Disinfect Staging
HD-DEN-16680_P1B Inlet Water Shutdown During Heat Disinfect Staging
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Once it goes into prime pause state -> say, user removes the clamping and resume the priming -> again this state executes to check the clamping condition. Now, venous pressure is in range and raw f...

Once it goes into prime pause state -> say, user removes the clamping and resume the priming -> again this state executes to check the clamping condition. Now, venous pressure is in range and raw flow rate more than minimum, we get into else state with last set flag ( primeDialysateOcclusionDetected) and change the priming state back to blood priming initial states.

Alarm above has stop property. So triggering alarm will take us out of this state and go to Prime Pause state. So this else will never happen I think.

Alarm above has stop property. So triggering alarm will take us out of this state and go to Prime Pause state. So this else will never happen I think.

This looks like a threshold. So name should start with MIN or MAX.

This looks like a threshold. So name should start with MIN or MAX.

HD-DENBUG-249_Pressure Window Starting Values P1 Staging
HD-DENBUG-249_Pressure Window Starting Values P1 Staging
HD-DENBUG-239_Alarm 160 Blood Pump Rotor SP
HD-DENBUG-239_Alarm 160 Blood Pump Rotor SP