DEN-14497, DEN-14326 Added a function to regularly schedule the informative blood leak embedded mode commands. Changed the FALSE to ACk not required in some of the message ack functions

  1. … 3 more files in changeset.
DEN-14344: addressed code review comments

  1. … 1 more file in changeset.
Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

  1. … 1 more file in changeset.
DEN-14344: removed venous high limit increase while air pump is operating. Increased air pump operation timeout to 6s to handle worst case bp ramping.

DEN-14344: changed air pump upper level persistence from 2s to 100ms.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-14007 merged staging and resolved the merge conflicts

  1. … 1 more file in changeset.
DEN-14007, DEN-14099 Merge & fix conflicts.

  1. … 1 more file in changeset.
DEN-14344: code review fixes.

DEN-14344: code review fixes.

Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

  1. … 2 more files in changeset.
DEN-14344: code review fixes.

  1. … 1 more file in changeset.
DEN-14344: code review comments

Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

  1. … 7 more files in changeset.
DEN-14344: compiler fixes

DEN-14344: fixed merge conflict missing lines.

Merge branch 'staging' into DEN-14344-sw-s85-mg-air-pump

  1. … 6 more files in changeset.
DEN-14344: Misc formatting fixes.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

  1. … 6 more files in changeset.
DEN-14344: fixes for adjusted venous limits when air pump is on

DEN-14307: partial fix to DEN-14086.

DEN-14344: complier fixes and warnings

  1. … 1 more file in changeset.
DEN-14344: created air pump monitor function for alarm id 316. Created a timer for venous pressure to handle adjusted treatment values for a time while the air pump is operating. Allow HD to start filling after consumable testing in pretreatment instead of waiting after dry self tests. formatted ifndef release code blocks to look neater.

  1. … 1 more file in changeset.
DEN-14007 Code Review comments update 1.

  1. … 1 more file in changeset.
DEN-14344: implmented venous pressure limits to increase when air pump is on. Reworked Air trap logic to turn off after air in the upper sensor for 2 seconds. Changed air pump timeout to be 5s.

DEN-14307 updated the code to get the heat disinfect temperature sensors (THd) and checks it value prior to starting a treatment

  1. … 2 more files in changeset.
DEN-14307: HD rotor vs. motor check - changed RPM threshold from 5 to 2 RPM.

DEN-14344: implementation with of alarm id 316