DEN-13460 updated the power loss alarm

Merged DEN-13460

  1. … 5 more files in changeset.
DEN-13460: Restored B term for blood flow estimation back t0 0.00055.

Merge branch 'DEN-13427-sw_dev-sprint-75_mg' into develop

  1. … 3 more files in changeset.
DEN-13460: Removed obsolete fixed PBA build switch.

  1. … 1 more file in changeset.
DEN-13427: Reworked DG's Service mode handling

  1. … 3 more files in changeset.
DEN-13460: Removed unused occlusion offset override support.

  1. … 3 more files in changeset.
DEN-13460: Added venous occlusion alarm.

  1. … 3 more files in changeset.
Merged DEN-13460

  1. … 5 more files in changeset.
DEN-13460 updated the code

  1. … 4 more files in changeset.
DEN-13460 added the voltages persistent alarm

  1. … 1 more file in changeset.
DEN-13460: Added handling of request to re-send HD alarms. Expanded CAN message buffer sizes.

  1. … 6 more files in changeset.
DEN-13460: Add minimum syringe pump rate as criteria for syringe pump direction alarm.

DEN-13460: Use max time windowed count instead of persistence for commutation errors.

DEN-13460: Added minimum pump speed as criteria for pump direction alarms.

Updated the valves transition algorithm

DEN-13460 updated the pressure occlusion

  1. … 1 more file in changeset.
Updated the syringe pump retract function

  1. … 1 more file in changeset.
Merged DEN-13460 and resolved the conflicts

  1. … 1 more file in changeset.
Updated the drivers with the latest demo changes

  1. … 1 more file in changeset.
DEN-13460: Fixed syringe home issues.

  1. … 1 more file in changeset.
DEN-13460 updated the valves homing code

  1. … 2 more files in changeset.
DEN-13460: Relaxed alarm 159 threshold. Reverted back to old way of estimating implied motor speed from target flow for purposes of alarm 159.

DEN-13460: Fixed merge conflicts.

DEN-13460: Added persistence to pump direction error count alarm. Fixed comments and pos A average calc for valves.

  1. … 2 more files in changeset.
Merged DEN-13460

DEN-13460 updated the code for DVT

  1. … 2 more files in changeset.
DEN-13460: Fixed bug in syringe pump POST test.

DEN-13460: Baseline HD pump direction error counters after POST completes so we don't alarm due to power-up.

DEN-13460 added a config around the syringe pump self test