Merged DEN-11750 and resolved the merge conflicts

  1. … 3 more files in changeset.
DEN-11750 updated the blood leak driver to send Ctrl U multiple times and make sure the sensor is communicating

  1. … 9 more files in changeset.
Changed the BLD self test timeout

DEN-11250 Merged branch DEN-11750 dg_dev-dg dialysate temperature tune up to branch DEN-11250 hd syringe pump update before working on syringe pump bug ticket #11824

  1. … 1 more file in changeset.
DEN-11250 Implemented new requirements for syringe pump occlusion & empty detection - when prime is done, wait 1 sec, calc delta force sensor greater than 0.5then trigger occlusion alarm

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

  1. … 2 more files in changeset.
DEN-11750: Added build switch to use PBo connected sensor as PBA and fix PBo to 100 mmHg.

  1. … 1 more file in changeset.
DEN-11750: Updated blood flow estimate constants.

  1. … 1 more file in changeset.
DEN-11250 Moved startHeprinBolus to Wet Self Tests

  1. … 4 more files in changeset.
Merged DEN-11750 and resolved the merge conflicts

  1. … 3 more files in changeset.
DEN-11750 removed the new blood leak zeroing method

  1. … 3 more files in changeset.
Merged DEN-11750 and resolved merge conflicts

  1. … 5 more files in changeset.
DEN-11750 added fans duty cycle override

  1. … 8 more files in changeset.
Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

DEN-11750: changed approach to saving reset reason code before startup code clears it.

  1. … 2 more files in changeset.
Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

  1. … 1 more file in changeset.
DEN-11750 work in progress. Worked on the BLD

  1. … 3 more files in changeset.
DEN-11750: Fixed handling of insufficient Heparin alarm (DEN-11824).

  1. … 1 more file in changeset.
Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

  1. … 1 more file in changeset.
DEN-11750: Fixed bug DEN-11823.

DEN-11750: Fixed alarm recover bug DEN-11824 for insufficient Heparin alarm. Fixed bug DEN-9692 to allow wider inline pressure range while ramping up blood pump.

  1. … 4 more files in changeset.
DEN-11750: Fixed bug where alarm 91 occurs if you turn on HD before DG.

DEN-11250 Fixed syringe pump direction alarm 79 triggers when attempting to seek

DEN-11816 Fixed syringe pump direction alarm 79 triggers when attempting to seek

  1. … 1 more file in changeset.
Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

DEN-11750 removed the DEBUG_DENALI

Fixed merge conflicts.

  1. … 5 more files in changeset.
DEN-11750: Fixed issue with syringe pump running while off alarm persistence.

  1. … 1 more file in changeset.
DEN-11750 updated the reservoirs and blood leak drivers

  1. … 1 more file in changeset.
Added blood leak ignore alarm

  1. … 1 more file in changeset.