DENBUG-104: Fixed errors in bug fix.

DEN-16691: Limit arterial pressure to >= -200 mmHg for Qb estimation purposes.

DEN-16691: Limit arterial pressure to >= -200 mmHg for Qb estimation purposes.

DENBUG-104: restored max dialysate pump speed to 4000 RPM. Apply reduced lower rotor speed tolerance only to commanded speeds > 3200 RPM.

DENBUG-104: restored max dialysate pump speed to 4000 RPM. Apply reduced lower rotor speed tolerance only to commanded speeds > 3200 RPM.

DENBUG-104: Increased max dialysate pump speed from 3400 to 3600 RPM. Reduced lower rotor speed tolerance from -5 RPM or -15% (whichever is bigger) to -60%.

DENBUG-104: Increased max dialysate pump speed from 3400 to 3600 RPM. Reduced lower rotor speed tolerance from -5 RPM or -15% (whichever is bigger) to -60%.

DENBUG-163-2-raw air trap level test support

  1. … 3 more files in changeset.
DENBUG-104: Allow for dialin pwm set cmd to go to full speed.

DENBUG-104: Allow for dialin pwm set cmd to go to full speed.

Fixed merge conflict from DENBUG-104.

  1. … 2 more files in changeset.
DENBUG-104: Cap dialysate pump speed to 3,400 RPM to prevent rotor speed errors.

DENBUG-104: Cap dialysate pump speed to 3,400 RPM to prevent rotor speed errors.

DEN-16672 resolving merge issues

DEN-16672 update

DEN-16672 vraiable airpump stop time based on bloodflow

DEN-16672 merge into develop

DEN-16672 stop air pump update

DEN-16672 resolving merge issues

DEN-16672 stop air pump after defined time

DEN-16672 stop air pump after defined time

DEN-16672 resolving merge issues

DEN-16672 airpump on delayed after fill events

resolving merge issues

  1. … 2 more files in changeset.
DEN-16672-using raw level for airpump off and reduced persistence

resolving merge issues

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
DENBUG-163 code review comments

DENBUG-163-raw air trap level sensor changes

  1. … 3 more files in changeset.
DEN-16672 resolving merge issues