ModeInitPOST.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

There is supposed to be 2 blank lines above and below this banner.

There is supposed to be 2 blank lines above and below this banner.

DEN-17220 added the code from staging

  1. … 7 more files in changeset.
HD-DEN-17220_Release 20 P1B Staging
HD-DEN-17220_Release 20 P1B Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DENBUG-294 addressed the code review comment

Done

Done

Not needed. Remove.

Not needed. Remove.

This is sent only once

This is sent only once

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DENBUG-294 addressed the code review comments

  1. … 1 more file in changeset.
Moved it to the top.

Moved it to the top.

Do we not have a serial # type already defined somewhere? I don't like defining types locally within a function.

Do we not have a serial # type already defined somewhere? I don't like defining types locally within a function.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Why are we doing this here? And why repeatedly every 500ms? Why not just once after NV POST state?

Why are we doing this here? And why repeatedly every 500ms? Why not just once after NV POST state?

This has been deferred for release 21.

This has been deferred for release 21.

DENBUG-294 added code to send the serial number and institutional record in POST

  1. … 1 more file in changeset.
Why aren't we treating blood leak recovering alarm the same as creating dialysate alarm?

Why aren't we treating blood leak recovering alarm the same as creating dialysate alarm?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

I think "dialysate" should be "dialyzer".

I think "dialysate" should be "dialyzer".

HD-DENBUG-290_Blood Leak Alarm During Rinseback Staging
HD-DENBUG-290_Blood Leak Alarm During Rinseback Staging