DEN-12224: Explicitly set actuators on pre-treatment transition to functions. Testing longer reservoir settle time.

  1. … 2 more files in changeset.
DEN-12224 Fixed DEN-12346 HD prime out of time alarm [ 227] Systems agreed to increase MAX_PRIME_TIME from 10 min to 15 minutes

DEN-12224 merged staging and added the switches monitor softwar configuration

  1. … 35 more files in changeset.
DEN-12224: Slowed blood side prime rates to 150 mL/min maximum.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 43 more files in changeset.
DEN-12215 Merged Staging branch. Updated DG_FILL_ defs, WET_SELF_TESTS_ defs.

  1. … 51 more files in changeset.
DEN-12224 updated the software configurations

  1. … 2 more files in changeset.
DEN-12224 updated the software configurations and increased the wet self test tolerances

  1. … 2 more files in changeset.
DEN-12224 added F and the end of the defines the are floating numbers

  1. … 24 more files in changeset.
DEN-12224 added _RELEASE_ to all the software configurations enums

  1. … 14 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 46 more files in changeset.
DEN-11750 merged staging and resolved the merge conflicts

  1. … 36 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 59 more files in changeset.
DEN-11750 updated the files with the software confiugrations

  1. … 21 more files in changeset.
DEN-11750 updated the files with software configurations

  1. … 14 more files in changeset.
DEN-11750 added software configurations for the primes

  1. … 20 more files in changeset.
DEN-11750 updated the fans moitor and added temperatures and fans POST states

  1. … 10 more files in changeset.
Disabled the software configurations

  1. … 6 more files in changeset.
DEN-11750 updated the blood leak driver to send Ctrl U multiple times and make sure the sensor is communicating

  1. … 10 more files in changeset.
DEN-11750: Changed blood prime and rinseback target volumes to be based on selected dialyzer type.

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-11098 Resolved merge conflicts

  1. … 5 more files in changeset.
DEN-11098 Removed code related to run blood circuit again because it no longer needed

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 112 more files in changeset.
DEN-11098 Addressed code review comments

  1. … 2 more files in changeset.
Disabled venous pressure check. Increased prime max time to 15 min.

  1. … 1 more file in changeset.
DEN-11098 Updated prime time limit for the saline dialyzer state

DEN-11098 Converted from seconds to milliseconds fix

DEN-11098 Corrected state transition and changed blood pump speed from 400 to 300 per system specification

DEN-11098 Incorporated peer review comments

  1. … 1 more file in changeset.