Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2

DEN-11114 initial commit of the reservoirs module. Continued with implementing the reservoir management

  1. … 5 more files in changeset.
DEN-11114: Updated UF control to included estimated UF flow as input.

  1. … 3 more files in changeset.
DEN-11089 Incorporated peer review comments

Done

Done

Fixed

Fixed

I believe reservoir 2 is used. static PRE_TREATMENT_RESERVOIR_MGMT_STATE_T handlePreTreatmentReservoirMgmtStartState( void ) { PRE_TREATMENT_RESERVOIR_MGMT_STATE_T state = PRE_TREATMENT_RESERVOIR_...

I believe reservoir 2 is used.

static PRE_TREATMENT_RESERVOIR_MGMT_STATE_T handlePreTreatmentReservoirMgmtStartState( void )
{
PRE_TREATMENT_RESERVOIR_MGMT_STATE_T state = PRE_TREATMENT_RESERVOIR_MGMT_START_STATE;

if ( TRUE == fillReservoirOneStartRequested )

Unknown macro: { fillReservoirOneStartRequested = FALSE; state = PRE_TREATMENT_RESERVOIR_MGMT_DRAIN_CMD_STATE; cmdSetDGActiveReservoir( DG_RESERVOIR_2 ); }


return state;
}

Fixed

Fixed

Done

Done

Done

Done

This might be reservoir 1 - there is a pre-treatment reservoir mgmt function handling this - check to see which reservoir is active at beginning of prime here.

This might be reservoir 1 - there is a pre-treatment reservoir mgmt function handling this - check to see which reservoir is active at beginning of prime here.

Why reset start time here?

Why reset start time here?

I think this needs to be closed or saline will not be forced across dialyzer into dialysate flow path.

I think this needs to be closed or saline will not be forced across dialyzer into dialysate flow path.

Make it clearer that this is for prime dialyzer state.

Make it clearer that this is for prime dialyzer state.

Dialyzer is spelled with a "z" not "s".

Dialyzer is spelled with a "z" not "s".

I would rename to something like prime dialyzer stop weight.

I would rename to something like prime dialyzer stop weight.

DEN-11089 if priming for the 1st time, transition to the HD_PRIME_SALINE_SETUP_STATE

DEN-11089 In hadlePrimeDialysateBypassState, state should be HD_PRIME_RESERVOIR_ONE_FILL_COMPLETE_STATE

DEN-11089 Swapped the dialysate priming circuit - prime the R2 circuit first then prime the R1 circuit

DEN-11089 Implemented new saline dialyzer priming state to purge air from the dialyzer path

HD-DEN-11098_Implement HD Pre Treatment Mode Update
HD-DEN-11098_Implement HD Pre Treatment Mode Update
DEN-11114: Reset BP rotor count before each treatment starts. Allow BP rotor count override to work even if WORN_CARTRIDGE build flag is enabled.

  1. … 1 more file in changeset.
Disabled watchdog and safety shutdown POST tests for unit 2.

  1. … 10 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 57 more files in changeset.
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH