Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

  1. … 1 more file in changeset.
DEN-11750: Fixed handling of insufficient Heparin alarm (DEN-11824).

  1. … 2 more files in changeset.
Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

  1. … 5 more files in changeset.
DEN-11750: Fixed alarm recover bug DEN-11824 for insufficient Heparin alarm. Fixed bug DEN-9692 to allow wider inline pressure range while ramping up blood pump.

  1. … 8 more files in changeset.
Fixed merge conflicts.

  1. … 12 more files in changeset.
DEN-11750: Changed saline bolus rate from 150 to 300 mL/min per Dec 8 2021 systems mtg.

DEN-11750: Changed blood prime and rinseback target volumes to be based on selected dialyzer type.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 35 more files in changeset.
Merged DEN-11114 and resolved the merge conflicts

  1. … 11 more files in changeset.
DEN-11098 Fixed merge conflicts

  1. … 35 more files in changeset.
DEN-11114 merged staging

  1. … 35 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-11098 Resolved merge conflicts

  1. … 4 more files in changeset.
DEN-11098 Removed code related to run blood circuit again because it no longer needed

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 98 more files in changeset.
Merge branch 'DEN-10602_UI_DEV_Add_PatientID_PreTreatment' into develop

DEN-11098 Addressed code review comments

  1. … 2 more files in changeset.
DEN-11114, DEN-11364 worked on the blood leak calibration and removed the fans clear the alarm if the RPM was in range again

  1. … 5 more files in changeset.
Fixed error in last.

Disabled entire BLD module. No t/o on no cartridge self-tests.

  1. … 1 more file in changeset.
Disabled venous pressure check. Increased prime max time to 15 min.

  1. … 1 more file in changeset.
Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

  1. … 1 more file in changeset.
Relaxed pressure recovery dry test requirement.

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

  1. … 2 more files in changeset.
DEN-1111: Fix dry pressure test. Fix wet flow vs. res test. Adjust test code patch for PBA in BP flow estimator.

  1. … 2 more files in changeset.
Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

  1. … 1 more file in changeset.
DEN-11098 Updated prime time limit for the saline dialyzer state

DEN-11114: Fixed AC POST test.

Merged DEN-11114 and resolved merge conflict