DEN-7605: Added safety shutdown state to alarm info broadcast.

  1. … 1 more file in changeset.
DEN-7860: Initial push for HD air bubble detectors module

  1. … 2 more files in changeset.
DEN-8103: Cleaned up TODO and added alarms

  1. … 17 more files in changeset.
DEN-8030 added override functions for the HD valves

  1. … 2 more files in changeset.
Merge branch 'DEN-7347-dg-hd_dev-post-treatment' into staging

DEN-7605: Moved alarm data type structure definition to common AlarmDefs.h. Extended syringe pump off persistence from 500ms to 1100ms due to 1 sec speed check interval.

  1. … 1 more file in changeset.
DEN-7347: Addressed code review

  1. … 2 more files in changeset.
DEN-7347: Addressed code reviews

  1. … 5 more files in changeset.
DEN-7395: Addressed merge conflicts

  1. … 9 more files in changeset.
DEN-7605: Added Dialin overrides to alarm info.

  1. … 5 more files in changeset.
DEN-7605: Fixed typo.

  1. … 2 more files in changeset.
DEN-7395: Implemented state machine architecture for Blood Leak Detector driver

  1. … 1 more file in changeset.
DEN-7605: Added no minimize flag to alarm status message to UI.

  1. … 3 more files in changeset.
Merge branch 'staging' into DEN-7091-pre-treatment-mode

  1. … 20 more files in changeset.
DEN-5980 resolved merge conflicts

  1. … 10 more files in changeset.
DEN-7347: Integrated syringe pump driver

  1. … 8 more files in changeset.
DEN-7091: Integrate syringe pump driver

  1. … 7 more files in changeset.
DEN-5980 addressed code review comments and merged staging

  1. … 12 more files in changeset.
DEN-7117: Added checks and alarms.

  1. … 4 more files in changeset.
DEN-6372: Merge from staging pt 2.

  1. … 15 more files in changeset.
DEN-6372: merged from staging.

  1. … 13 more files in changeset.
DEN-7091: Polling alarm stop status

  1. … 8 more files in changeset.
DEN-7091: Added handler for UI sample command request

  1. … 6 more files in changeset.
DEN-6372: Fixes per lab testing. Added function to AlarmMgmt to determine whether any alarms w/ stop property are active. Changed treatment mode handling of alarm stop from signal to modes approach to modes polling alarm mgmt approach. Added f/w support for override of treatment time remaining.

  1. … 7 more files in changeset.
DEN-5980 added comm messages to be able to set and get NV records

  1. … 14 more files in changeset.
DEN-6402: Added DG valve setting change command

  1. … 7 more files in changeset.
DEN-6402: Stored command response for each possible command

  1. … 4 more files in changeset.
DEN-6402: Added alarm signal action handlers

  1. … 7 more files in changeset.
DEN-6372: Amended treatment stop sub-mode.

  1. … 6 more files in changeset.
DEN-6402: Added wet self-tests state machine