Services

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'DEN-7605-hd_dg_dev-self-tests' into develop

DEN-7605: minor fixes.

Merge branch 'DEN-8103-dg-hd_dev-alarm-design' into develop

  1. … 10 more files in changeset.
Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

  1. … 1 more file in changeset.
Merged DEN-8030

  1. … 4 more files in changeset.
.

  1. … 3 more files in changeset.
DEN-8030 updated the valves controller

  1. … 4 more files in changeset.
DEN-7605: Added safety shutdown state to alarm info broadcast.

Fixed merge conflicts.

  1. … 2 more files in changeset.
DEN-7605: Added build switches.

  1. … 4 more files in changeset.
Need to remove this comment.

Need to remove this comment.

Need to remove this comment.

Need to remove this comment.

Same as comment from DialinFlow.c

Same as comment from DialinFlow.c

We can't have 3 calls to isPersistentAlarmTriggered() on same alarm. If the first call starts the timer and the second or third call is good, it will reset the timer to 0 again. Suggest creating a ...

We can't have 3 calls to isPersistentAlarmTriggered() on same alarm. If the first call starts the timer and the second or third call is good, it will reset the timer to 0 again.
Suggest creating a BOOL variable to have all three conditions or with each other and call isPersistentAlarmTriggered() only once.

DEN-7860: Initial push for HD air bubble detectors module

  1. … 2 more files in changeset.
DEN-7605: minor fixes to compatibility checks.

  1. … 1 more file in changeset.
DEN-7605: Added FPGA compatibility to FPGA POST test. Added skeleton POST for f/w compatibility.

  1. … 1 more file in changeset.
initBloodPrime() is called every time we transition to blood prime sub-mode.

initBloodPrime() is called every time we transition to blood prime sub-mode.

Blood prime last motor count does not seem to get updated except in initBloodPrime(). Should it be updated continuously somewhere?

Blood prime last motor count does not seem to get updated except in initBloodPrime(). Should it be updated continuously somewhere?

DEN-8103: Cleaned up TODO and added alarms

  1. … 14 more files in changeset.
DEN-7605: Added FPGA read timeout checks/alarms for blood/dialysate flow.

  1. … 2 more files in changeset.
Fixed merge conflicts.

  1. … 2 more files in changeset.
DEN-7605: Added FPGA pressure sensor checks and alarms.

  1. … 1 more file in changeset.
DEN-8103: Updated fpga comm time out alarm name

DEN-7605: Added FPGA reported comm error to FPGA comm error check. Fixed persistent alarm periods in syringe pump module.

  1. … 4 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.