RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Need to verify SRS and AlarmsDef.h are in sync.

Need to verify SRS and AlarmsDef.h are in sync.

Alignment

Alignment

Comments line up but the values don't?

Comments line up but the values don't?

Done.

Done.

Done.

Done.

Extra white space.

Extra white space.

Should align

Should align

Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Updated to 600.

Updated to 600.

So R2 fill was 1100 and you say it ended up at 500. Would be 800 if keep same at 1100. If we change to 600, would end around 300 as you say - I think that's right.

So R2 fill was 1100 and you say it ended up at 500. Would be 800 if keep same at 1100. If we change to 600, would end around 300 as you say - I think that's right.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Old transfer was 600mL, new is 300mL. Res 1 should end up with 1500 so added 300 to baseline, Res 2 was previously ending with 500. Do we want it to end R2 with 300?

Old transfer was 600mL, new is 300mL. Res 1 should end up with 1500 so added 300 to baseline, Res 2 was previously ending with 500. Do we want it to end R2 with 300?

Updated for only reservoir 2 fill state.

Updated for only reservoir 2 fill state.

There are two dialysate prime states. First goes through dialyzer and both DPs are running and we still want 300 rate. Second bypasses dialyzer and only DPi is running and we want 600 rate.

There are two dialysate prime states. First goes through dialyzer and both DPs are running and we still want 300 rate. Second bypasses dialyzer and only DPi is running and we want 600 rate.

This looks like we're skipping draining of reservoirs before doing any filling if priming disabled. I think we always want to drain both reservoirs first. Also, if we go to fill state like this, wi...

This looks like we're skipping draining of reservoirs before doing any filling if priming disabled. I think we always want to drain both reservoirs first.
Also, if we go to fill state like this, will it do a flush or a normal fill? I think it will do a flush fill since flushed flag is still FALSE.

Remove extra blank line.

Remove extra blank line.

We should initialize these with a loop iterating through each reservoir (in case a reservoir is added/removed at some point).

We should initialize these with a loop iterating through each reservoir (in case a reservoir is added/removed at some point).

This should probably be 600 since we'll only be transferring 300 to R1 in wet self-tests.

This should probably be 600 since we'll only be transferring 300 to R1 in wet self-tests.

Alignment

Alignment

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Yes, may need to be done when merging to Staging. Is this flag used independently anywhere? Or planned to be?

Yes, may need to be done when merging to Staging. Is this flag used independently anywhere? Or planned to be?

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

I don't think I have this bug fix in my branch. Needs to be addressed in a more recent branch.

I don't think I have this bug fix in my branch. Needs to be addressed in a more recent branch.

Good catch. Updated Voltages.c to behave same as Battery.c (trigger the correction alarm depending on whether in a treatment mode).

Good catch. Updated Voltages.c to behave same as Battery.c (trigger the correction alarm depending on whether in a treatment mode).

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

This check for _IN_TREATMENT may need to be added to Valves.c ALARM_ID_HD_AC_POWER_LOST check from recent bug fix. Both look to be checked in every case. Is this flag used independently anywhere?

This check for _IN_TREATMENT may need to be added to Valves.c ALARM_ID_HD_AC_POWER_LOST check from recent bug fix. Both look to be checked in every case. Is this flag used independently anywhere?