HD-DENBUG-158_2 Alarm 128 Venous Air Bubble Triggered Again During Post TX Sequence
HD-DENBUG-158_2 Alarm 128 Venous Air Bubble Triggered Again During Post TX Sequence
HD-DENBUG-186_Alarm 68 Dialysate Gen Recirculation Error Clears Alarm 147 Dialysate Gen
HD-DENBUG-186_Alarm 68 Dialysate Gen Recirculation Error Clears Alarm 147 Dialysate Gen
HD-DENBUG-198_2 Alarm 130 142 Triggered And The End OF A Rinseback End OF Treatment
HD-DENBUG-198_2 Alarm 130 142 Triggered And The End OF A Rinseback End OF Treatment
HD-DENBUG-158_Alarm 128 Venous Air Bubble Triggered Again During Post TX Sequence
HD-DENBUG-158_Alarm 128 Venous Air Bubble Triggered Again During Post TX Sequence
HD-DENBUG-186_Alarm 68 Dialysate Gen Recirculation Error Clears Alarm 147 Dialysate Gen
HD-DENBUG-186_Alarm 68 Dialysate Gen Recirculation Error Clears Alarm 147 Dialysate Gen
HD-DENBUG-160_Alarm 147 Dialysate Gen Not Ready Cleared With Alarm 97 And Did Not Appear ON UI
HD-DENBUG-160_Alarm 147 Dialysate Gen Not Ready Cleared With Alarm 97 And Did Not Appear ON UI
HD-DENBUG-116_DG Allows Update OF Calibration Service System Records IN Other Modes Besides
HD-DENBUG-116_DG Allows Update OF Calibration Service System Records IN Other Modes Besides
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

The result will be an unsigned integer

The result will be an unsigned integer

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

In the If conditions, we check "uFVolume > colUFVol" conditions, that makes uFRate always positive and hence the condition for minimum check is not required.