DEN-11114: perform periodic clock check only if release build so debugging will not cause alarm.

Done.

Done.

DEN-11089 Removed signalDialOutPumpHardStop since it has not been started

You can remove this line here. That pump has not yet been run.

You can remove this line here. That pump has not yet been run.

DEN-11114: Removed obsolete UF PI controller. Cleaned up UF control code changes.

DEN-11089 the VDO pump is stopped before transition from dialyzer state to wet selftests state

Fixed. Before transition from dialyzer state to wet selftests state, the VDO pump should be stopped.

Fixed. Before transition from dialyzer state to wet selftests state, the VDO pump should be stopped.

Move this line to when prime dialyzer state is completed before we go to wet self tests.

Move this line to when prime dialyzer state is completed before we go to wet self tests.

DEN-11089 Correction - dialyzer dialysate bypass priming path uses R1 instead of R2 and dialyzer dialysate priming path uses R2 instead of R1

DEN-11114: Added comment.

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2

DEN-11114 initial commit of the reservoirs module. Continued with implementing the reservoir management

    • -0
    • +407
    ./Services/Reservoirs.c
    • -0
    • +38
    ./Services/Reservoirs.h
DEN-11114: Updated UF control to included estimated UF flow as input.

DEN-11089 Incorporated peer review comments

Done

Done

Fixed

Fixed

I believe reservoir 2 is used. static PRE_TREATMENT_RESERVOIR_MGMT_STATE_T handlePreTreatmentReservoirMgmtStartState( void ) { PRE_TREATMENT_RESERVOIR_MGMT_STATE_T state = PRE_TREATMENT_RESERVOIR_...

I believe reservoir 2 is used.

static PRE_TREATMENT_RESERVOIR_MGMT_STATE_T handlePreTreatmentReservoirMgmtStartState( void )
{
PRE_TREATMENT_RESERVOIR_MGMT_STATE_T state = PRE_TREATMENT_RESERVOIR_MGMT_START_STATE;

if ( TRUE == fillReservoirOneStartRequested )

Unknown macro: { fillReservoirOneStartRequested = FALSE; state = PRE_TREATMENT_RESERVOIR_MGMT_DRAIN_CMD_STATE; cmdSetDGActiveReservoir( DG_RESERVOIR_2 ); }


return state;
}

Fixed

Fixed

Done

Done

Done

Done

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

DEN-11114: Restored previous motor controller speed feedback conversion to RPM - it was right before.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

This might be reservoir 1 - there is a pre-treatment reservoir mgmt function handling this - check to see which reservoir is active at beginning of prime here.

This might be reservoir 1 - there is a pre-treatment reservoir mgmt function handling this - check to see which reservoir is active at beginning of prime here.

Why reset start time here?

Why reset start time here?

I think this needs to be closed or saline will not be forced across dialyzer into dialysate flow path.

I think this needs to be closed or saline will not be forced across dialyzer into dialysate flow path.

Make it clearer that this is for prime dialyzer state.

Make it clearer that this is for prime dialyzer state.

Dialyzer is spelled with a "z" not "s".

Dialyzer is spelled with a "z" not "s".