DEN-11098 Updated code per peer review comments

Fixed.

Fixed.

Replaced by primeDialysateDialyzerTimeLimit because this variable is dynamic and based on the type of dialyzer.

Replaced by primeDialysateDialyzerTimeLimit because this variable is dynamic and based on the type of dialyzer.

Add blank line between these two lines.

Add blank line between these two lines.

Add blank line between these lines.

Add blank line between these lines.

Do we need separate start time variables? These are not used at same time - consider consolidating into a single variable.

Do we need separate start time variables? These are not used at same time - consider consolidating into a single variable.

Should be 120 mL divided by 300 mL/min times 60 sec/min = 24 seconds.

Should be 120 mL divided by 300 mL/min times 60 sec/min = 24 seconds.

Why is this commented out?

Why is this commented out?

DEN-11098 Updated time out limits

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

DEN-11114, DEN-11372 added the timer reset to prevent error 168 alarm

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-10602_UI_DEV_Add_PatientID_PreTreatment' into develop

DEN-11098 Implemented code update due to requirements change

HD-DEN-10602_UI Dev Add Patient ID Pre Treatment
HD-DEN-10602_UI Dev Add Patient ID Pre Treatment
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Enabled and disabled alarms

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

DEN-11114 worked on the reservoirs and heaters control

DEN-11114: Added correction factor and updated constants for blood flow estimation.

Finishing up the story for firmware

DEN-10607: Implementation [Screen]

DEN-10609: Implementation [Treatment Log]

DEN-11114: Remove obsolete BP speed vs. flow alarm.

DEN-11098 Corrected delta volume calculationbased on the past 10 seconds value instead of the past 9 seconds

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Fixed DG_RESERVOIR to DG_RESERVOIRS

Merge branch 'DEN-10153_Treatment_Log_Prescribed_UF' into staging

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo