HD-DENBUG-162_Low Venous Pressure Alarm Triggers During Normal Treatment Operation
HD-DENBUG-162_Low Venous Pressure Alarm Triggers During Normal Treatment Operation
Done.

Done.

Done.

Done.

MIN_TIME_BETWEEN_AIR_TRAP_FILL_LOW_PRESSURE_ALARM_EXEMPTIONS_MS (as a suggestion). Also be more descriptive in comments.

MIN_TIME_BETWEEN_AIR_TRAP_FILL_LOW_PRESSURE_ALARM_EXEMPTIONS_MS (as a suggestion).
Also be more descriptive in comments.

Add OR condition for if useShort == FALSE

Add OR condition for if useShort == FALSE

Use u32DiffWithWrap() function in utilities.

Use u32DiffWithWrap() function in utilities.

This looks wrong.

This looks wrong.

HD-DENBUG-162_Low Venous Pressure Alarm Triggers During Normal Treatment Operation
HD-DENBUG-162_Low Venous Pressure Alarm Triggers During Normal Treatment Operation
HD-DEN-16499_FW Message List Testing Issues
HD-DEN-16499_FW Message List Testing Issues
the second level 'if' statement indentation is appeared to be missing for readability.

the second level 'if' statement indentation is appeared to be missing for readability.

HD-DEN-15953_Dialysate Temp Alarm Triggered After Dialysate Recirculation Stopped
HD-DEN-15953_Dialysate Temp Alarm Triggered After Dialysate Recirculation Stopped
HD-DEN-16563_Venous Pressure Readings Were Below 20Mmhg For 1.5 Minutes Before Low Venous
HD-DEN-16563_Venous Pressure Readings Were Below 20Mmhg For 1.5 Minutes Before Low Venous
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Do we want to have the 11% minimum here? I thought it was just for closed loop control.

Do we want to have the 11% minimum here? I thought it was just for closed loop control.

This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit?

This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit?

HD-DEN-16519_Pressure Window Adjustments DO Not Reset After New Treatment
HD-DEN-16519_Pressure Window Adjustments DO Not Reset After New Treatment
HD-DEN-16519_Pressure Window Adjustments DO Not Reset After New Treatment
HD-DEN-16519_Pressure Window Adjustments DO Not Reset After New Treatment