Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-11750: Fixed handling of insufficient Heparin alarm (DEN-11824).

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Enabled syringe pump, syringe alarms and inline pressure alarms.

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

DEN-11750: Fixed bug DEN-11823.

DEN-11750: Fixed alarm recover bug DEN-11824 for insufficient Heparin alarm. Fixed bug DEN-9692 to allow wider inline pressure range while ramping up blood pump.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Updated the build switches

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Enabled HD pump alarms for Tom.

DEN-11750: Fixed bug where alarm 91 occurs if you turn on HD before DG.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Enabled syringe alarms.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-11250 Fixed syringe pump direction alarm 79 triggers when attempting to seek

DEN-11816 Fixed syringe pump direction alarm 79 triggers when attempting to seek

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Disabled prime build switches

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Disabled and enabled build switches

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

DEN-11750 removed the DEBUG_DENALI

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Fixed merge conflicts.

DEN-11750: Fixed issue with syringe pump running while off alarm persistence.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Enabled syringe pump and related alarms for Eliza.

DEN-11750 updated the reservoirs and blood leak drivers

DEN-11750: Changed saline bolus rate from 150 to 300 mL/min per Dec 8 2021 systems mtg.