Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-6200: Reduced fill volume during priming

DEN-6372: Minor fix to saline bolus.

DEN-6200: Added pumps, heaters, and valves settings for Pre-Treatment Recirculate state

DEN-6200: Updated persistent alarms

DEN-6372: Fix to PI controller anti-windup logic.

DEN-6200: Change timer counter to use timer lib

DEN-6200: Added time out alarm for priming dialyzer and bypass circuit

DEN-6200: Added prime complete alarm to pre-treatment mode

DEN-6200: Removed unused functions and variables

DEN-5674: Resolved merge conflicts from staging.

  1. … 4 more files in changeset.
Merge branch 'staging' into DEN-5887-hd-ultrafiltration-filtering

DEN-5674: Addressed code review comments.

DEN-5674: Addressed code review comments.

DEN-5674: Addressed code review comments.

DEN-5674: Addressed code review feedback.

DEN-5674: Addressed code review feedback.

DEN-5674: Addressed code review feedback.

DEN-5887: Addressed code review feedback. Removed obsolete alarm features.

DEN-6200: Added dialyzer and bypass fluid path priming

DEN-5674: Make all code comments start with a capital letter for consistency.

    • -104
    • +104
    ./Controllers/DialInFlow.c
  1. … 42 more files in changeset.
DEN-5674: Addressed code review comments.

  1. … 13 more files in changeset.
DEN-6200: Home valves in pre-treatment no cartridge self-test

DEN-5381: Increase purge air time out to 30 seconds

DEN-5674: Added msg ID for auto-ACK to enum. Misc. comments. Removed dead code.

DEN-5674: Changed size of alarm silence cmd from U08 to U32 for consistency.

DEN-5381: Added transition to prime and removed ui workflow build switch

DEN-5381: Wait for blood pump finished homing before transition to prime

DEN-5381: Reset pre-treatment state machine and deleted unused code

DEN-5381: Moved DG start and valves homing into pre-treatment mode

DEN-5674: Fixed alarm audio volume control per HDD change.