DEN-12550 updated the switches driver

DEN-12224 fixed the switches enable high/low

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-12224, DEN-12536 updated the post treatment mode

DEN-12224, DEN-12536 updated the post treatment's reservoir states

DEN-12224, DEN-12550 reversed the open/close status of the pump track switch

DEN-12224, DEN-12536 updated post treatment mode to drain both reservoirs

Dara removed this line of code after lab test w/o this change could not replicate the bug.

Dara removed this line of code after lab test w/o this change could not replicate the bug.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-12224

DEN-12224 updated the consumables self test

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Disabled code for testing

Why is this necessary. This is already called somewhere else.

Why is this necessary. This is already called somewhere else.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

DEN-12224: fixed reservoir post-fill settle time bug.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

DEN-12224: .

DEN-12224: fixed merge conflicts.

DEN-12224: Explicitly set actuators on pre-treatment transition to functions. Testing longer reservoir settle time.

DEN-12224 updated the code

DEN-12224 updated the code

DEN-12224 Fixed DEN-12346 HD prime out of time alarm [ 227] Systems agreed to increase MAX_PRIME_TIME from 10 min to 15 minutes

DEN-12224 Fixed ticket DEN-12533 HD should accepting only value for arterial pressure high from -270 to 0

DEN-12224 Added build switch to disable syringe pump removed alarm[219] on system without syringe pump installed