HD-DEN-16372_Srsdg 907 Not Met
HD-DEN-16372_Srsdg 907 Not Met
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

We should not be doing a Heparin bolus in treatment mode. It should be completed in pre-treatment mode. transitionToDialysis() function will resume Heparin dispense (not bolus) if necessary when tr...

We should not be doing a Heparin bolus in treatment mode. It should be completed in pre-treatment mode. transitionToDialysis() function will resume Heparin dispense (not bolus) if necessary when treatment is resumed.

Are these F32s?

Are these F32s?

HD-DENBUG-84_Heparin Bolus Does Not Resume After Treatment Pause
HD-DENBUG-84_Heparin Bolus Does Not Resume After Treatment Pause
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

HD-DEN-16090_2 Sys Saline Bubbles Enter Pbo Line During Cartridge Priming
HD-DEN-16090_2 Sys Saline Bubbles Enter Pbo Line During Cartridge Priming
HD-DENBUG-83_Alarm 278 Triggered Active Res Weight Out OF Range
HD-DENBUG-83_Alarm 278 Triggered Active Res Weight Out OF Range
HD-DENBUG-78_Dvt 21 Alarm 110 Triggers AT Start OF Treatment With Heparin ON
HD-DENBUG-78_Dvt 21 Alarm 110 Triggers AT Start OF Treatment With Heparin ON