BloodFlow.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-736: Tuned blood flow control coefficients.

  1. … 4 more files in changeset.
LDT-736: updated BP stroke volume.

LDT-736: Added basic flow estimation calculation - still needs Leahi volume per rotor rev value from Systems. Added BP homing logic to handle issue with FPGA always reporting home until it has seen a home to not home transition.

  1. … 2 more files in changeset.
LDT-736 updated the code for UI demo

  1. … 9 more files in changeset.
LDT-736: Implemented BP home operation per latest FPGA hall sensor design.

  1. … 6 more files in changeset.
LDT-736: Added support for Dialin request to home the BP. Added test build switches to disable specific checks that aren't working on alpha 3 unit.

  1. … 6 more files in changeset.
LDT-736: Moved air pump Dialin set cmd from driver to controller. Fixed ADC channel overrun. Allow multiple ejector homes.

  1. … 7 more files in changeset.
LDT-469: Removed HET edge interrupt - no longer needed for rotor hall sensors. Updated BP rotor speed function.

  1. … 11 more files in changeset.
LDT-94: Added basic treatment mode, dialysis state, treatment parameters mode. Added skeletal treatment paused state.

  1. … 28 more files in changeset.
LEAH-838: Addressed code review feedback.

  1. … 17 more files in changeset.
LEAH-838: Fix blood pump stop command. Fix raw level sensor readings. Added air trap level control test code build switch (temporary code for special test build).

  1. … 5 more files in changeset.
LEAH1970: Refactored component codes to new style of codes for consistency.

  1. … 26 more files in changeset.
TEST: refactor component codes to new codes for consistency.

  1. … 26 more files in changeset.
LEAH-1970: Increased BP ramp rate. Misc cosmetic fixes.

TEST: misc cosmetic fixes.

TEST: Changed BP ramp rate to 1000 RPM per second.

TEST: changes during alpha prototype integration.

  1. … 14 more files in changeset.
TEST: compensate commanded BP speed so that it will match measured speed.

  1. … 3 more files in changeset.
TEST: Fixed blood speed feedback issue - coming back as a period instead of RPM. Fixed pressure conversion.

  1. … 5 more files in changeset.
LEAH-39: Fixed include issue.

  1. … 1 more file in changeset.
LEAH-51: Addressed code review comments.

  1. … 1 more file in changeset.
LEAH-51: Further developed BloodFlow unit. Addressed compiler errors/warnings.

  1. … 6 more files in changeset.
LEAH-51: Addressed code review comments.

  1. … 2 more files in changeset.
Added air trap, level sensors, valves.

  1. … 31 more files in changeset.